Model
|
CS-151F1
|
CS-152F1
|
CS-150F1
|
Target of
measurement
|
SC-1
(ammonia/hydrogen peroxide
aqueous solution)
|
SC-2
(hydrochloric acid/hydrogen
peroxide aqueous solution)
|
SPM
(sulfuric acid/hydrogen
peroxide aqueous solution)
|
Multi-monitoring with one unit of up to 4 types of chemical solutions or ranges possible.
|
Measurement principle
|
Absorption spectrometry
|
Concentration computation
principle
|
Temperature-compensated multivariate analysis
|
Measurement range
(mass %)
|
NH3: 0.00% to 1.00%
H2O2: 0.00% to 5.00%
H2O: 94.0% to 100.0%
|
HCI: 0.00% to 2.00%
H2O2: 0.00% to 2.00%
H2O: 96.0% to 100.0%
|
H2SO4: 70.0% to 96.0%
H2O2: 0.00% to 10.00%
H2O: 4.0% to 30.0%
|
Repeatability (mass %)
(within ±1% room temperature)
|
NH3: ±0.15%
H2O2: ±0.30%
H2O: ±1.5%
|
HCI: ±0.15%
H2O2: ±0.15 %
H2O: ±1.5%
|
H2SO4: ±0.50%
H2O2: ±0.50%
H2O: ±1.5%
|
・Repeatability for the list above is for the cell unit type.Please contact us for the repeatability of the in-line cell type.
・Differences in the above list may occur depending on measurement range.(Please see our website for details.)
|
Sample solution temperature
|
20 to 80: (when using Inline cell) 20 to 30: (when using cell unit)
|
Measurement cycle
|
Approx. 3 seconds
|
Content of
LCD and input
|
• 1st component: measured concentration value (%) •Error code
• 2nd component: measured concentration value (%) •Parallel concentration alarm output invalid status
• 3rd component: measured concentration value (%) •Serial/parallel status
• 1st and 2nd components: concentration alarm status •Background compensation data
|
RS-232C serial input
|
• Serial/parallel status switching •Parallel concentration alarm output status
• Measurement data transfer request (valid/invalid)
• 1st and 2nd component concentration alarm setting •Target measurement solution switching
|
Parallel input
|
12 to 30 V DC (photocoupler insulated) • Parallel concentration alarm output
• Target measurement solution switching
• Background compensation
|
Parallel output
|
Open collector output (photocoupler insulated) • 1st component concentration alarm, 4 types
Max. current when on: • 2nd component concentration alarm, 4 types
DC 5 mA (no internal protection resistor) • Device error
Max. impressed voltage when off: 30 V DC • Measuring
• Target measurement solution type
• Background compensation status
|
Solution temperature input
|
Terminal block connection (M3 terminal screws)
• Platinum resistance temperature sensor: Pt100 (Pt100 sensor should be prepared by customers)
|
Analog output
|
• 1st component concentration alarm, 4 types: 4 to 20 mA
• 2nd component concentration alarm, 4 types: 4 to 20 mA
✽We also support other measurement ranges. Please specify when ordering.
|
Fiber optic cable
|
Dedicated CS-100F1 Series fiber optic cable (25 ±15: installation temperature)
|
Sample cell
|
We can supply dedicated sample cells (inline cell or cell unit) for various kinds of chemical solutions.
|
Power supply
|
100 V to 230 V AC (single phase) ±10%, 50/60 Hz, approx. 100 VA
|
Dimensions
|
205 (W) x 329 (D) x 269 (H) mm 8.1 (W) x 13.0 (D) x 10.6 (H) in (excluding protruding part)
|
Weight
|
Approx. 11 kg
|