The MA886 is an optical instrument that employs the measurement of the refractive index to determine sodium chloride concentration in aqueous solutions used in food preparation. It is not intended for sea water salinity measurements. The measurement of refractive index is simple and quick and provides the user an accepted method for NaCl analysis. Samples are measured after a simple user calibration with deionized or distilled water. Within seconds the instrument measures the refractive index of the solution. The digital refractometer eliminates the uncertainty associated with mechanical refractometers and is easily portable for measurements where you need them.
The instrument utilizes internationally recognized references for unit conversion and temperature compensation. It can display the measurement of NaCl concentration 4 different ways: g/100 g g/100 mL Specific Gravity and °Baumé. Temperature (in °C or °F) is displayed simultaneously with the measurement (on 3 of the ranges) on the large dual level display along with icons for Low Power and other helpful message codes.
Key features include: • Waterproof models offers IP65 waterproof protection • Automatic Temperature Compensation (ATC) • Battery operation with Low Power indicator • Automatically turns off after 3 minutes of non-use
技术参数:
SPECIFICATIONS Range
0 to 28 g/100 g 0 to 34 g/100 ml 1.000 to 1.216 Specific Gravity 0 to 26 ° Baumé 0 to 80°C / 32 to 175°F
Resolution
0.1 g/100 g 0.1 g/100 ml 0.001 Specific Gravity 0.1 ° Baumé 0.1°C / 0.1°F
Accuracy ±0.2 g/100 g ±0.2 g/100 ml ±0.002 Specific Gravity ±0.2 ° Baumé ±0.3°C / ±0.5°F Light Source Yellow LED Measurements Time Approximately 1.5 seconds Minimum Sample Volume 100 µL (cover prism totally) Sample Cell SS ring and flint glass prism Temperature Compensation Automatic between 10 and 40°C / 50 to 104°FCase Material ABS Enclosure Rating IP 65 Battery Type 1 x 9V AA (included) Battery Life 5000 reading Auto-shut off after 3 minutes of non-use Dimensions 192 x 102 x 67 mm Weight 420 g
产品描述:
It is not intended for sea water salinity measurements. The measurement of refractive index is simple and quick and provides the user an accepted method for NaCl analysis. Samples are measured after a simple user calibration with deionized or distilled water. Within seconds the instrument measures the refractive index of the solution. The digital refractometer eliminates the uncertainty associated with mechanical refractometers and is easily portable for measurements where you need them.
Temperature (in °C or °F) is displayed simultaneously with the measurement (on 3 of the ranges) on the large dual level display along with icons for Low Power and other helpful message codes.
• Waterproof models offers IP65 waterproof protection
• Automatic Temperature Compensation (ATC)
• Battery operation with Low Power indicator
• Automatically turns off after 3 minutes of non-use
0 to 34 g/100 ml
1.000 to 1.216 Specific Gravity
0 to 26 ° Baumé
0 to 80°C / 32 to 175°F
0.1 g/100 ml
0.001 Specific Gravity
0.1 ° Baumé
0.1°C / 0.1°F
±0.2 g/100 ml
±0.002 Specific Gravity
±0.2 ° Baumé
±0.3°C / ±0.5°F Light Source Yellow LED Measurements Time Approximately 1.5 seconds Minimum Sample Volume 100 µL (cover prism totally) Sample Cell SS ring and flint glass prism Temperature Compensation Automatic between 10 and 40°C / 50 to 104°FCase Material ABS Enclosure Rating IP 65 Battery Type 1 x 9V AA (included) Battery Life 5000 reading Auto-shut off after 3 minutes of non-use Dimensions 192 x 102 x 67 mm Weight 420 g